UV-106-W miernik ozonu w wodzie View larger

Ozone meter in water UV-106-W

New product

The UV-106-W meter is one of the most accurate and universal devices for measuring ozone in water.

Product features:

Application - Measurement of ozone concentration in clean and dirty water

Ozone concentration range 0-100 ppm (g / m3, μg / mL)

Accuracy Above 0.05 ppm or 1% of value

Measurement frequency 10 s

Response time 20 s

Averaging times of 10 s, 1 min, 5 min, 1 hr

Display units of concentration ppm, μg / ml, mg / l, g / m3

The waiting time for UV-106-W is about 4 weeks from the date of payment.

More details

  • The ozone concentration meter in water UV-106-W uses our patented MicroSparge ™ technology which is used to measure the concentration of ozone dissolved in water with high accuracy. Unlike most such devices, our equipment does not use a membrane that will wear out over time. Instead, the ozone concentration is measured by almost completely spraying about 2 ml of water containing ambient air with ozone, and then filtered to obtain only ozone particles in the gas state taken from the sample. The result obtained is subjected to a slight correction based on the temporary ozone profile removed from the solution, so as not to miss any particles that may have remained in the solution before filtering. Due to the fact that the concentration of ozone is measured in a gaseous state, any interference caused by other molecules and dissolved organic and inorganic chemical compounds are removed, which makes it possible to use the device in both perfectly clean and polluted water, such as drinking water, which it may contain many different dissolved organic and inorganic impurities as well as suspended particles.

    Picture

    Measurement method (absolute method)  

    Integrated ozone absorbance with UV radiation completely extracted from a 2 ml water sample  

     

    Application  

    Measurement of ozone concentration in clean and dirty water  

     

    Ozone concentration range  

    0-100 ppm (g / m 3 , μg / mL)  

     

    Accuracy  

    Above 0.05 ppm or 1% of reading  

     

    Drift zero

    <0.05 ppm per month

     

    Measurement frequency

    10 p

     

    Response time

    20 p

     

     

    Averaging times of measurements

     

    10 s, 1 min, 5 min, 1 hr

     


    Displayed concentration units

    ppm, μ g / ml, mg / l, g / m 3

     

     

    Energy demand

     

    11-14 V DC, 2.2 A at 12 V, 26.4 W

     

     

    Average water sample flow rate

     

    300 cm 3 / min

     

    Pressure range

    0-50 psi (> 100 psi instantaneous pressure)

     

    Housing

    NEMA

     

    Relays (2 in additional power tester)

    Resolution 0.1 ppm, 2 levels, voltage free SPDT contacts

     

    Analogue outputs

    4-20 mA, 0-2.5 V (2-point scalable)

     

    Digital outputs

    LCD, RS232, USB

     

     

    Journal record

    Internal data log, 16,368 lines (10-sec average =

    1.9 days; 5-min average = 57 days)

    LED warnings

    Low lamp, Low flow, Incorrect measurement

     

    dimensions

    13.3 x 12.0 x 7.3 in (33.8 x 30.5 x 18.5 cm)

     

    Weight

    14.5 lb (6.6 kg)